FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1914544 · Received December 3, 2010

Report

Report Number
6000001-2010-05442
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 10, 2010
Report Date
November 10, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF ?OCCLUSION, UPSTREAM, FALSE ALARM?. (B)(4)

Additional Manufacturer Narrative · 1

THE CONDITION OF CONSTANT FALSE UPSTREAM OCCLUSION ALARMS ON CHANNEL A WAS CONFIRMED THROUGH EVALUATION DUE TO A FAULTY PHM (PUMP HEAD MODULE). THE CHANNEL A PHM WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)

Description of Event or Problem · 1

DURING DEVICE EVALUATION BY BAXTER ON A COLLEAGUE INFUSION PUMP, CONSTANT FALSE UPSTREAM OCCLUSION ALARMS OCCURRED ON CHANNEL A. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.04.00 CATEGORIZED AS UNREMEDIATED.