Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION:BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF FAILURE CODE 550.320.844.0000. (B)(4)

Additional Manufacturer Narrative · 1

THE CONDITION OF FAILURE CODE 550.320.844.0000 THAT DID NOT AUDIBLY ALARM WAS CONFIRMED THROUGH EVALUATION. THE CONDITION OCCURRED DUE TO A DISCONNECTED REAR HARNESS. THE REAR HARNESS WAS RE-CONNECTED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4)

Description of Event or Problem · 1

DURING DEVICE EVALUATION BY BAXTER A FAILURE CODE 550.320.844.0000 OCCURRED WITH NO AUDIBLE ALARM ON A COLLEAGUE INFUSION PUMP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.04.00 CATEGORIZED AS UNREMEDIATED.