SPEEDSCREW 5.5 IMPLANT WITH INSERTER HANDLE
Report
- Report Number
- 2032380-2010-00027
- Event Type
- Other
- Date Received
- November 23, 2010
- Date of Event
- October 22, 2010
- Report Date
- November 23, 2010
- Manufacturer
- ARTHROCARE CORP.
- Product Code
- MBI
- PMA / PMN Number
- K081893
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PT INFO WAS REQUESTED BUT WAS NOT AVAILABLE. A LOT HISTORY REVIEW WAS PERFORMED FOR THE DEVICE LOT. NO ABNORMALITIES WERE IDENTIFIED THAT WOULD LEAD TO THE REPORTED PRODUCT PROBLEM. THE DEVICE WAS RECEIVED UNDEPLOYED. THE IMPLANT WAS FOUND INTACT. THE SNARE LINES WERE FOUND WRAPPED AROUND THE REELS WITHOUT CLINICAL SUTURE. THE GREEN SNARE LINE WAS FOUND INTACT AND THE WHITE SNARE LINE WAS FOUND WITH ITS LOOP BROKEN OFF. THE SNARE LOOP WAS FOUND ON THE DISTAL END OF THE SHAFT. IT IS POSSIBLE THAT THE LOOP BROKE DUE TO EXCESSIVE TENSION APPLIED DURING THE CINCHING PROCESS; HOWEVER, AN EXACT ROOT CAUSE IS NOT KNOWN. A SECOND DEVICE USED IN THE SAME PROCEDURE WAS FILED UNDER MDR 2032380-2010-00026. (B)(4).
THE PT UNDERWENT A ROTATOR CUFF REPAIR AND SUBACROMIAL DECOMPRESSION SURGERY USING TWO OPUS SPEEDSCREW 5.5 IMPLANTS. FOR THE FIRST IMPLANT USED, THE DEVICE FAILED TO TENSION DURING THE LOADING OF THE SUTURE INTO THE IMPLANT AND THE IMPLANT WAS REMOVED AND SECOND IMPLANT PLACED IN THE SAME BONE HOLE. DURING THE TENSION OF THE SECOND IMPLANT, THE CROSS PIN BROKE AND THE SURGEON COULD NOT USE THE DEVICE AS PROPER TENSIONING WAS NOT ACHIEVED. THE SURGEON ELECTED TO MOVE TO A MINI-OPEN PROCEDURE TO COMPLETE THE REPAIR USING A BONE TUNNEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPEEDSCREW 5.5 IMPLANT WITH INSERTER HANDLE | FASTENER, FIXATION, NONDEGRADABLE, SOFT | MBI | ARTHROCARE CORP. | 1011605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |