FDA Adverse Event Malfunction Summary report: N

RESPIRONICS

MDR report key: 19144215 · Received April 19, 2024

Report

Report Number
2518422-2024-20975
Event Type
Malfunction
Date Received
April 19, 2024
Date of Event
April 2, 2024
Manufacturer
RESPIRONICS, INC.
Product Code
MNT
UDI-DI
00884838020054
PMA / PMN Number
K102985
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: G1 PHONE NUMBER +1(724)351-2041

Additional Manufacturer Narrative · 0

THE CUSTOMER INFORMED THE REMOTE SERVICE ENGINEER (RSE) OF THE ISSUE AND REQUESTED THE PART NUMBERS REQUIRED FOR A REPAIR. THE RSE PROVIDED THE CUSTOMER WITH THE PART INFORMATION FOR THE 2ND GENERATION LIQUID CRYSTAL DISPLAY (LCD), 2ND GENERATION LCD CABLE, 2ND GENERATION USER INTERFACE (UI) PRINTED CIRCUIT BOARD ASSEMBLY (PCBA), LCD TO UI PCBA CABLE, 2ND GENERATION LCD TRAY, AND SCREW PACK. MULTIPLE GOOD FAITH EFFORTS (GFE) WERE ATTEMPTED TO OBTAIN CONFIRMATION OF THE PART ORDER, PART REPLACEMENT, AND RESOLUTION. NO RESPONSE OR FURTHER INFORMATION WAS RECEIVED FROM THE CUSTOMER. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 0

PHILIPS RECEIVED A COMPLAINT ON THE V60 VENTILATOR INDICATING THAT THE DISPLAY WAS NOT WORKING. THE DEVICE WAS REPORTED TO BE OUTSIDE OF USE AT THE TIME OF THE REPORTED PROBLEM. NO PATIENT OR USER HARM REPORTED. THE CUSTOMER INFORMED THE REMOTE SERVICE ENGINEER (RSE) OF THE ISSUE AND REQUESTED THE PART NUMBERS REQUIRED FOR A REPAIR. THE RSE PROVIDED THE CUSTOMER WITH THE PART INFORMATION FOR THE 2ND GENERATION LIQUID CRYSTAL DISPLAY (LCD), 2ND GENERATION LCD CABLE, 2ND GENERATION USER INTERFACE (UI) PRINTED CIRCUIT BOARD ASSEMBLY (PCBA), LCD TO UI PCBA CABLE, 2ND GENERATION LCD TRAY, AND SCREW PACK. THIS INVESTIGATION IS ONGOING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1695651 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS, INC. V60 V60PLUS VENTILATOR 00884838020054

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown