IMP,TSV,4.7,11.5,MTX,MG
Report
- Report Number
- 0002023141-2024-01228
- Event Type
- Injury
- Date Received
- April 19, 2024
- Date of Event
- October 22, 2023
- Report Date
- September 11, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024020009
- PMA / PMN Number
- K101977
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A4: WEIGHT UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K101880. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. ZIMVIE RECEIVED THE REPORTED IMPLANT FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE. IMPLANT FRACTURE IDENTIFIED AT THE COLLAR, AND A BROKEN SCREW WAS FOUND INSIDE OF IT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1252891. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1252891 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE EXCESSIVE OCCLUSAL FORCES - CUSTOMER ERROR. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE PATIENT RETURNED TO OUR OFFICE ON REFERRAL FROM THEIR RESTORING DENTIST WITH A LOOSE IMPLANT CROWN. ON EXAM, THE NECK OF THE IMPLANT AT TOOTH LOCATION #14 WAS CRACKED. THE IMPLANT WAS REMOVED AND THE SOCKET GRAFTED FOR FUTURE IMPLANT RE-PLACEMENT.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1694639 | IMP,TSV,4.7,11.5,MTX,MG | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1252891 | 00889024020009 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Male | Required Intervention |