FDA Adverse Event
Other
Summary report: N
HORIZONTALDIAGNOST H
MDR report key: 1914366
·
Received November 18, 2010
Report
- Report Number
- 3003768251-2010-00204
- Event Type
- Other
- Date Received
- November 18, 2010
- Date of Event
- March 11, 2008
- Report Date
- March 11, 2008
- Manufacturer
- PHILIPS MEDICAL SYSTEMS DMC GMBH
- Product Code
- IXR
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE SYSTEM IS APPROX 30 YEARS OLD AND THE COLLIMATOR LAMP TIMER BOARD WAS REDESIGNED DUE TO COMPONENT OBSOLESCENCE. THE SYSTEM HAD BEEN UPDATED WITH THE NEW TIMER BOARD AND THE CUSTOMER FELT IT WAS TOO LONG AND THE LAMP COVER COULD GET TOO HOT. HOWEVER, THE FIXED SETTING OF 45 SECONDS IS PER PRODUCT SPECIFICATION. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER ALLEGED THAT THE COVER ON THE COLLIMATOR LAMP GOT TOO HOT WITH A FIXED LAMP TIMER SETTING OF 45 SECONDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HORIZONTALDIAGNOST H | IXR (RADIOGRAPHIC TABLE) | IXR | PHILIPS MEDICAL SYSTEMS DMC GMBH | 70404 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |