FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAS PRDGM INS V2.2 SK EN
MDR report key: 1914090
·
Received November 19, 2010
Report
- Report Number
- 3004209178-2010-83411
- Event Type
- Malfunction
- Date Received
- November 19, 2010
- Date of Event
- October 26, 2010
- Report Date
- October 26, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND AND THE DEVICE FAILED THE DISPLACEMENT TEST AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER HAD MULTIPLE MOTOR ERROR ALARMS FOR THE PAST TWO WEEKS. ADVISED THE CUSTOMER TO DISCONTINUE USED OF THE INSULIN PUMP AND TO REVERT TO BACK UP PLAN OF MANUAL INJECTIONS. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |