FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1914090 · Received November 19, 2010

Report

Report Number
3004209178-2010-83411
Event Type
Malfunction
Date Received
November 19, 2010
Date of Event
October 26, 2010
Report Date
October 26, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND AND THE DEVICE FAILED THE DISPLACEMENT TEST AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD MULTIPLE MOTOR ERROR ALARMS FOR THE PAST TWO WEEKS. ADVISED THE CUSTOMER TO DISCONTINUE USED OF THE INSULIN PUMP AND TO REVERT TO BACK UP PLAN OF MANUAL INJECTIONS. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR