SITTER SELECT
Report
- Report Number
- 2020362-2010-00367
- Event Type
- Malfunction
- Date Received
- November 19, 2010
- Report Date
- October 4, 2010
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4), RESULTS: EVALUATION OF THE PRODUCT FOUND THAT THERE WAS ONE BATTERY SPRING COMPLETELY COVERED BY BATTERY ACID THEREFORE, NO FURTHER TESTS COULD BE PERFORMED. THERE IS NO VISIBLE DAMAGE TO THE ALARM CASE. NOTE: THE INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: DO NOT MIX OLD AND NEW BATTERIES OR BATTERY BRANDS. THIS MAY CAUSE RUPTURE OR LEAKAGE AND DAMAGE ALARM. (B)(4).
THE CUSTOMER REPORTED THAT THE ALARM HAS NO POWER WHEN TESTED WITH NEW BATTERIES ALL OF THE SAME BRAND AND WITH NEW SENSOR PADS. THERE WAS NO DAMAGE DETECTED BY THE CUSTOMER ON THE ALARM CASE. THIS WAS DISCOVERED DURING SET UP PRIOR TO PLACING IT INTO USE WITH A PATIENT. INSPECTION OF THE PRODUCT FOUND THAT ONE OF THE BATTERY SPRINGS INSIDE THE BATTERY COMPARTMENT IS COMPLETELY COVERED BY BATTERY ACID.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SITTER SELECT | KMI | J. T. POSEY CO. | 8361 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |