FDA Adverse Event Injury Summary report: N

NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM

MDR report key: 1914006 · Received December 3, 2010

Report

Report Number
6000034-2010-00757
Event Type
Injury
Date Received
December 3, 2010
Date of Event
November 19, 2010
Report Date
April 14, 2011
Manufacturer
COCHLEAR LTD.
Product Code
MCM
PMA / PMN Number
970051
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
AUDIOLOGIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PER THE CLINIC, AT INITIAL ACTIVATION THE PATIENT HAD NO REPORTED AUDITORY PERCEPTS. UPON FURTHER INVESTIGATION IT WAS FOUND THAT THE ELECTRODE ARRAY WAS NOT IN THE COCHLEA. THE DEVICE WAS EXPLANTED ON (B)(6), 2010 AND THE PATIENT WAS REIMPLANTED WITH ANOTHER DEVICE DURING THE SAME SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM MCM, PRODUCT CODE: MCM MCM COCHLEAR LTD. CI512 N/A

Patients

Seq Age Sex Outcome Treatment
1 35 MO Required Intervention