FDA Adverse Event
Injury
Summary report: N
NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM
MDR report key: 1914006
·
Received December 3, 2010
Report
- Report Number
- 6000034-2010-00757
- Event Type
- Injury
- Date Received
- December 3, 2010
- Date of Event
- November 19, 2010
- Report Date
- April 14, 2011
- Manufacturer
- COCHLEAR LTD.
- Product Code
- MCM
- PMA / PMN Number
- 970051
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- AUDIOLOGIST
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
PER THE CLINIC, AT INITIAL ACTIVATION THE PATIENT HAD NO REPORTED AUDITORY PERCEPTS. UPON FURTHER INVESTIGATION IT WAS FOUND THAT THE ELECTRODE ARRAY WAS NOT IN THE COCHLEA. THE DEVICE WAS EXPLANTED ON (B)(6), 2010 AND THE PATIENT WAS REIMPLANTED WITH ANOTHER DEVICE DURING THE SAME SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM | MCM, PRODUCT CODE: MCM | MCM | COCHLEAR LTD. | CI512 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 MO | Required Intervention |