FDA Adverse Event Malfunction Summary report: N

BONE SPREADER 12MM BEAK WIDTH

MDR report key: 19140041 · Received April 18, 2024

Report

Report Number
8030965-2024-05262
Event Type
Malfunction
Date Received
April 18, 2024
Manufacturer
SYNTHES GMBH
Product Code
HWN
UDI-DI
10886982202413
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: D4: LOT NUMBER ADDED. D9: DEVICE RETURNED. H4: MANUFACTURE DATE ADDED. SERVICE AND REPAIR EVALUATION: THE CUSTOMER REPORTED ¿IT WAS REPORTED THAT ON AN UNKNOWN DATE THE DEVICE WAS BROKEN, THERE WAS NO PATIENT OUTCOME.¿ THE REPAIR TECHNICIAN REPORTED THAT DEVICE FAILED COSMETICS TEST, THREAD SHAFT THREAD DAMAGE AND STOP NUT WAS MISSING. THE CAUSE OF THE ISSUE IS USER ERROR. THE ITEM WILL BE REPAIRED, PUT THROUGH FINAL INSPECTION, AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. FINALIZED SERVICE RECORD WILL BE ARCHIVED. THE EVALUATION WAS NOT CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE HISTORY REVIEW (DHR): PART # 399.13. SYNTHES LOT # 5901566. SUPPLIER LOT # 5901566. RELEASE TO WAREHOUSE DATE: 22 JUN 2009. SUPPLIER: SYNTHES GMBH. NO NCR'S GENERATED DURING PRODUCTION.. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT ON AN UNKNOWN DATE THE DEVICE WAS BROKEN. THERE WAS NO PATIENT OUTCOME. THIS REPORT IS FOR ONE (1) BONE SPREADER 12MM BEAK WIDTH THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2583268 BONE SPREADER 12MM BEAK WIDTH INSTRUMENT, COMPRESSION HWN SYNTHES GMBH 5901566 10886982202413

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown