FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1913883 · Received December 3, 2010

Report

Report Number
1423500-2010-06481
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 8, 2010
Report Date
November 8, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WILL BE PERFORMED FOR THE LOT NUMBER PROVIDED. THE PRODUCT SAMPLE HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED BY BAXTER FOR EVALUATION. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR A SYSTEM ERROR (SE) 2240 ALARM WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. BASED ON THE INFORMATION OBTAINED FROM BAXTER'S INVESTIGATION, THE ROOT CAUSE OF THIS REPORT WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A HOME PATIENT (HP) CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT MULTIPLE ALARMS THAT OCCURRED ON THE HOMECHOICE (HC) MACHINE DURING PRIMING. THE HP FURTHER STATED HE REVIEWED THE ALARM LOG AND NOTED SYSTEM ERROR (S/E) 2240 OCCURRED. THE HP STATED THAT HE USED THE SAME CASSETTE AFTER RELOADING IT AND THEN GOT REPEATED ALARMS. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED THE SET UP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. ON (B)(6) 2010 (B)(4) PRODUCT SURVEILLANCE CONTACTED THE HP WHO REPORTED NO FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H10H23102

Patients

Seq Age Sex Outcome Treatment
1