LIGAMAX-5MM
Report
- Report Number
- 3005075853-2010-06828
- Event Type
- Malfunction
- Date Received
- December 3, 2010
- Date of Event
- September 27, 2010
- Report Date
- September 28, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). JAWS DISENGAGED. THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED INSERTED THROUGH AN ENDOPATH XCEL TROCAR AND WITH ONE JAW DISENGAGED FROM THE CAM. THIS CONDITION WOULD NOT ALLOW THE JAWS TO COLLAPSE IN ORDER TO FORM THE CLIPS AND BE REMOVED FROM THE TROCAR. THE DEVICE WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE FOUND CONDITION. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE JAWS MAY BE INADVERTENT FORCE, TWISTING OR PRESSURE BEING PLACED ON THE DEVICE JAWS, USING THE JAWS OF THE DEVICE AS A DISSECTOR/RETRACTOR OR DAMAGE TO THE JAWS WHILE ENTERING THE TROCAR. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS OVER-TRAVELED. THIS FINDING IS NOT RELATED TO THE REPORTED EVENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE THE CLIPS WERE MALFORMED. WHEN THEY TRIED TO REMOVE THE DEVICE THE JAWS WOULD NOT CLOSE SO THEY REMOVED THE DEVICE AND THE TROCAR TOGETHER. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |