ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1913673
·
Received November 22, 2010
Report
- Report Number
- 2531779-2010-02486
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Report Date
- October 25, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE SWOLLEN, THE "UP", "DOWN" AND "OK" KEYPAD BUTTONS WERE UNRESPONSIVE TO USER INPUT, THE "CONTRAST" KEYPAD BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE "UP, DOWN, OK AND CONTRAST" KEY CONTACTS WERE INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE OK AND DOWN ARROW BUTTONS INTERMITTENTLY DID NOT RESPOND WHEN PRESSED. THE BUTTONS WOULD NOT SPRING BACK AFTER BEING PRESSED. IT WAS REPORTED, THERE WAS NO STRUCTURAL DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.