FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1913673 · Received November 22, 2010

Report

Report Number
2531779-2010-02486
Event Type
Malfunction
Date Received
November 22, 2010
Report Date
October 25, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE SWOLLEN, THE "UP", "DOWN" AND "OK" KEYPAD BUTTONS WERE UNRESPONSIVE TO USER INPUT, THE "CONTRAST" KEYPAD BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE "UP, DOWN, OK AND CONTRAST" KEY CONTACTS WERE INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE OK AND DOWN ARROW BUTTONS INTERMITTENTLY DID NOT RESPOND WHEN PRESSED. THE BUTTONS WOULD NOT SPRING BACK AFTER BEING PRESSED. IT WAS REPORTED, THERE WAS NO STRUCTURAL DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1