FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1913668 · Received November 22, 2010

Report

Report Number
2531779-2010-02477
Event Type
Malfunction
Date Received
November 22, 2010
Report Date
October 25, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE SWOLLEN, THE UP/DOWN/OK BUTTONS WERE INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE CONTRAST BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE UP KEY CONTACTS WAS MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED AND DID NOT ALWAYS SPRING BACK, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES WERE INTERMITTENTLY ACTIVATING THE DESIRED PUMP FUNCTIONS AND WERE STICKING. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO MOISTURE AND THERE IS NO STRUCTURAL DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.