FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1913626 · Received November 12, 2010

Report

Report Number
1720753-2010-04071
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 25, 2010
Report Date
November 12, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT UP. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPY XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1