FDA Adverse Event Malfunction Summary report: N

RESTORE RECHARGEABLE NEUROSTIMULATOR

MDR report key: 1913549 · Received November 12, 2010

Report

Report Number
3004209178-2010-09434
Event Type
Malfunction
Date Received
November 12, 2010
Date of Event
October 1, 2010
Report Date
October 28, 2010
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT EXPERIENCED COUPLING AND/OR COMMUNICATION ISSUES. THE PT FELL ON (B)(6) 2010, AND THE DEVICE DID NOT WORK AFTER THAT. THE PT DID NOT RECHARGE THE DEVICE IN THREE TO FOUR WEEKS PRIOR TO THIS REPORT. IT WAS SUSPECTED THAT AN OVER-DISCHARGE CONDITION EXISTED. THE STIMULATION WAS NOT ADJUSTED IN THE PREVIOUS THREE MONTHS. IT WAS SUGGESTED THAT THE INS HAD FLIPPED, BUT THIS WAS NOT CONFIRMED. IT WAS LATER REPORTED THAT THE PT WAS NOT ABLE TO ADJUST THE STIMULATION. A "CALL YOUR DOCTOR" ICON WAS DISPLAYED AND A POWER ON RESET (POR) CONDITION EXISTED. A WARNING POR WAS RECEIVED AFTER A PHYSICIAN MODE RECHARGE (PMR). THE PT WAS TO CONTACT THE MANUFACTURER REPRESENTATIVE TO SET UP AN APPOINTMENT TO CLEAR THE WARNING POR USING A MODEL 8840 PHYSICIAN PROGRAMMER. NO FURTHER DETAILS, PT SYMPTOMS OR OUTCOME WERE PROVIDED. ADDITIONAL INFO WAS REQUESTED BUT NOT RECEIVED AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFO BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE RECHARGEABLE NEUROSTIMULATOR LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37711 NA

Patients

Seq Age Sex Outcome Treatment
1 37 YR EXPLANTED:| IMPLANTED:| IMPLANTED:| EXTENSION: MODEL 37081, LOT # NJB004143N| IMPLANTED:| EXPLANTED:| IMPLANTED:| LEAD: MODEL 3777, LOT # J0546520V| EXPLANTED:| EXTENSION: MODEL 37081, LOT # NJB004142N| PROGRAMMER: MODEL 37742, LOT # NJD011824N| EXPLANTED:| EXPLANTED:| LEAD: MODEL 3777, LOT # N0029853| IMPLANTED: