FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1913461 · Received November 24, 2010

Report

Report Number
2531779-2010-02589
Event Type
Malfunction
Date Received
November 24, 2010
Report Date
October 29, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT (NO LIFTING OR PEELING OBSERVED), THE UP/DOWN/OK BUTTONS WERE INTERMITTENTLY RESPONDING DURING TESTING, THE CONTRAST BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE TESTING, THE CONTRAST KEY CONTACTS WERE INVERTED, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD PRESSES ARE INTERMITTENTLY ACTIVATING THE DESIRED PUMP FUNCTIONS. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO MOISTURE AND THERE IS NO STRUCTURAL DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1