FDA Adverse Event
Malfunction
Summary report: N
TERUMO CDI 500 BLOOD PARAMETER MONITOR
MDR report key: 1913429
·
Received November 12, 2010
Report
- Report Number
- 1828100-2010-02057
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 19, 2010
- Report Date
- November 12, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP
- Product Code
- DRY
- PMA / PMN Number
- K972962
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
EVALUATION IN PROGRESS, BUT NOT YET CONCLUDED.
Description of Event or Problem · 1
DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THE SERVICE TECHNICIAN REPORTED THE VENOUS BLOOD PARAMETER PROBE FAILED THE STANDARD REFERENCE SENSOR TEST. THE MONITOR WAS ORIGINALLY RETURNED FOR A CF08 ERROR MESSAGE. SINCE THE EVENT OCCURRED DURING ROUTINE TESTING OF THE DEVICE, THERE WAS NO PT INVOLVEMENT DURING THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO CDI 500 BLOOD PARAMETER MONITOR | ON LINE BLOOD GAS MONITOR | DRY | TERUMO CARDIOVASCULAR SYSTEMS CORP | 500AHCT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |