FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1913409 · Received November 24, 2010

Report

Report Number
2531779-2010-02583
Event Type
Malfunction
Date Received
November 24, 2010
Report Date
October 26, 2010
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT NO LIFTING OR PEELING OBSERVED. THE "UP/DOWN/OK" AND "CONTRAST" KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING DURING TESTING. THE "UP/DOWN" KEYPAD BUTTONS REQUIRE EXCESSIVE FORCE TO ACTIVATE DURING TESTING. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION, THE "UP" AND "CONTRAST" KEY CONTACTS ARE MISALIGNED. THE "UP/DOWN/CONTRAST" AND "OK" KEY CONTACTS BECOMES INVERTED WHEN PRESSED AND DO NOT ALWAYS SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE LAY-USER/REPORTER ALLEGED THAT THE OK BUTTON IS STICKING. DURING TROUBLESHOOTING, CUSTOMER SERVICE NOTED THE ONLY THE OK BUTTON WAS AFFECTED. THE PROBLEM ISSUE BEGAN WEEKS AGO. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS 2020 NA