PEN NDL 31GA 5MM 100BX 1200 USA

MDR report key: 19134076 · Received April 18, 2024

Report

Report Number
9616656-2024-05265
Event Type
Malfunction
Date Received
April 18, 2024
Report Date
July 11, 2024
Manufacturer
BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND
Product Code
FMI
UDI-DI
00382903201198
PMA / PMN Number
K213478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: D3 (COUNTRY TYPE AND COUNTRY), H6 (COMPONENT CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AS BENT CANNULA PE. BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.

Description of Event or Problem · 0

CONSUMER HAS DIFFICULTY REMOVING THE NEEDLE FROM THE PEN AFTER HIS INJECTION. HE STATED THAT WHEN HE TRIES TO PUT THE OUTER COVER BACK ONTO THE HUB, THE COVER DOES NOT FIT SECURELY AND IT FALLS OFF. CONSUMER DOES NOT RE-USE. BUT HE STORES NEEDLE ON HIS PEN AT TIMES. LOT #: 2130196. CATALOG #: 320119 DATE OF EVENT: UNKNOWN. SAMPLES: AWAITING SAMPLE. DC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2237717 PEN NDL 31GA 5MM 100BX 1200 USA NEEDLE, HYPODERMIC, SINGLE LUMEN FMI BECTON DICKNSON AND CO. - DUN LAOGHAIRE CO, IRELAND 320119 2130195 00382903201198

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown