- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCI FIELD SERVICE ENGINEER (FSE) VISITED (B)(6) 2010. THE CURRENT COMPRESSOR/RELAY WAS SHORTED. THE FSE REPLACED REFRIGERATION SLED ON (B)(6) 2010.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO SMOKE AND SMALL FLAMES OBSERVED FROM THE REFRIGERATION UNIT OF POWER PROCESSOR. NO OPERATOR WAS EXPOSED TO THE SMOKE AND THERE WAS NO EFFECT TO PATIENT OR USER.