UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-01479
- Event Type
- Malfunction
- Date Received
- December 2, 2010
- Date of Event
- November 3, 2010
- Report Date
- November 3, 2010
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
QC WAS ACCEPTABLE BEFORE THE EVENT, BUT QC ON CR-S HIGH CONTROL WAS A LITTLE LOW IN THE RANGE AFTER THE EVENT. A BCI CUSTOMER TECHNICAL SUPPORT (CTS) RECOMMENDED THE CUSTOMER TO USE PERSONAL PROTECTIVE EQUIPMENT (PPE) BEFORE TROUBLESHOOTING. THE CTS INSTRUCTED THE CUSTOMER TO FLUSH THE SAMPLE PROBE USING 10% WASH SOLUTION AND RINSE WITH DI WATER. THE CUSTOMER RERAN QC, AND ALL RESULTS WERE IN THE ESTABLISHED RANGES. THE CUSTOMER RAN THREE PATIENT SAMPLES ON BOTH INSTRUMENTS AND THE GLU, TBIL, CR-S, TG, AND UREA RESULTS MATCHED. THE CUSTOMER BELIEVES THE INSTRUMENT IS NOW WORKING CORRECTLY, BUT WILL CONTINUE MONITORING. AS OF (B)(4) 2010, THERE WERE NO FURTHER CALLS FOR THIS ISSUE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS MULTIPLE CHEMISTRY RESULTS GENERATED BY UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER FOR TWO PATIENTS. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER INDICATED THAT THERE WAS NO EFFECT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 600 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |