FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE

MDR report key: 19131749 · Received April 18, 2024

Report

Report Number
9610773-2024-30060
Event Type
Malfunction
Date Received
April 18, 2024
Date of Event
March 21, 2024
Report Date
January 13, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

G4: ADDITIONAL, K190744. THE EVALUATION OF THE EVENT IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE CORRECTION OF THE INITIAL REPORT AND THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE WAS NOT CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: SEVERE DENTS IN THE LIGHT GUIDE [FIBER] BUNDLE OF DISTAL END. BASED ON THE RESULTS OF THE INVESTIGATION, AND SINCE THE DEVICE MALFUNCTION WAS NOT CONFIRMED DURING EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE MALFUNCTIONED AND NOTICED THERE WAS ¿INSUFFICIENT BRIGHTNESS¿. THE ISSUE HAPPENED DURING PREPARATION PRIOR TO AN UNSPECIFIED THERAPEUTIC LAPAROSCOPIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
540522 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown OTV-S190.