PENTAX

MDR report key: 19131325 · Received April 18, 2024

Report

Report Number
9610877-2024-52941
Event Type
Malfunction
Date Received
April 18, 2024
Date of Event
April 10, 2024
Report Date
April 18, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
PMA / PMN Number
K172156
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB NOISE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE WITH DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE INSERTION FLEXIBLE TUBE COATING DAMAGE, AND THE REMOTE CONTROL BUTTONS WORN OUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2237387 PENTAX VIDEO NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL8-J10

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown