FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 1913121 · Received December 2, 2010

Report

Report Number
1628664-2010-00455
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
November 26, 2010
Report Date
November 30, 2010
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
LGS
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). TO EVALUATE THE CUSTOMER ISSUE, A REVIEW WAS CONDUCTED OF THE COMPLAINT TEXT, THE INSTRUMENT HISTORY, AND ARCHITECT SYSTEM LABELING. A DEFINITE CAUSE OF THE ERRATIC TOXO IGG RESULT WAS NOT FOUND. A REVIEW OF COMPLAINT TRACKING AND TRENDING METRICS WAS PERFORMED AND IDENTIFIED NO ADVERSE TRENDS IN CONJUNCTION WITH THE COMPLAINT ISSUE CURRENTLY UNDER EVALUATION. THE ARCHITECT SYSTEM OPERATIONS MANUAL ((B)(4)) AND THE ARCHITECT TOXO IGG REAGENT PACKAGE INSERT ((B)(4)) BOTH CONTAIN INFORMATION TO ADDRESS THE CUSTOMER'S CURRENT ISSUE. BASED ON THE AVAILABLE INFORMATION, A PRODUCT MALFUNCTION WAS NOT IDENTIFIED. AFTER REPROCESSING THE CONTROLS AND RECALIBRATING THE REAGENT, THE RESULTS WERE REPEATED WITHIN THE EXPECTED RANGES. REVIEW OF COMPLAINT TRACKING AND TRENDING; ABBOTT CUSTOMER TECHNICAL ADVOCATE INITIATED TROUBLESHOOTING WITH CUSTOMER INTERVENTION.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATES THAT A FALSE POSITIVE RESULT HAS BEEN GENERATED ON ONE SAMPLE FROM A NON-IMMUNIZED PREGNANT FEMALE PATIENT TESTED WITH THE ARCHITECT TOXO IGG ASSAY. THE SAMPLE GENERATED AN INITIAL RESULT OF 13 IU/ML THAT RETESTED AT 0.1 / 0.1 IU/ML. NO SUSPECT RESULTS WERE REPORTED FROM THE LAB. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER AUTOMATED IMMUNOASSAY ANALYZER LGS ABBOTT MANUFACTURING, INC.

Patients

Seq Age Sex Outcome Treatment
1 ARCH TOXO IGG LN: 6C19 LOT 89736HN00| ARCH TOXO IGG LN: 6C19 LOT 89736HN00