SYRINGE 1.0ML 30GA 1/2IN UF 10BAG 500CS
Report
- Report Number
- 1920898-2024-05100
- Event Type
- Malfunction
- Date Received
- April 18, 2024
- Date of Event
- January 23, 2024
- Report Date
- July 9, 2024
- Manufacturer
- BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949
- Product Code
- FMF
- UDI-DI
- 00382908411035
- PMA / PMN Number
- K190054
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
6 SYRINGES AFFECTED BY EVENT.
ADDITIONAL INFORMATION PROVIDED TO SECTIONS B4, G6, H2, H3, AND H11. SEE MEDWATCH COMPLETED SECTION C FORM ATTACHED. CORRECTIONS MADE TO SECTION H6 (COMPONENT CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, & INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
CONSUMER REPORTED FOUND 6 NEEDLE SHIELD LAYING IN 1 SEALED SYRINGE PACKET. 5 OF THE 6 HAD THE NEEDLES STUCK IN THE SHIELDS. AND 1 SYRINGE HAD THE NEEDLE ATTACHED TO THE SYRINGE WITH JUST THE EMPTY SHIELD IN THE BAG. NO INJURY FROM THIS INCIDENT. DC LOT # 3100603 CATALOG# 328411 DATE OF EVENT (B)(6) 2024 SAMPLE STATUS AWAITING SAMPLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 540483 | SYRINGE 1.0ML 30GA 1/2IN UF 10BAG 500CS | SYRINGE, PISTON | FMF | BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 | 328411 | 3100603 | 00382908411035 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |