FDA Adverse Event Other Summary report: N

SUNOPTIC TECHNOLOGIES

MDR report key: 1913083 · Received August 10, 2010

Report

Report Number
1035968-2010-00002
Event Type
Other
Date Received
August 10, 2010
Date of Event
June 23, 2010
Report Date
July 29, 2010
Manufacturer
SUNOPTIC TECHNOLOGIES C/O SERVICE AND TECH SUPPORT
Product Code
FSS
PMA / PMN Number
K961074
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

WITHOUT A PICTURE OR DETAILED DESCRIPTION OF THE ACTUAL BURN IT IS IMPOSSIBLE TO DETERMINE HOW THIS COULD HAVE OCCURRED. THE LIGHT SPOT FROM THE HEADLIGHT IS CIRCULAR AND IT WAS NOTED THAT THE BURN WAS "CRESANT SHAPED" WHICH CAUSES ONE TO WONDER HOW THE BURN COULD BE FROM THE HEADLIGHT. CERTAINLY ON A BREAST ONE MIGHT EXPECT THE SURFACE TISSUE TO BE SUCH THAT AN AREA RAPIDLY RECEDES FROM THE SOURCE OF THE LIGHT THUS REDUCING ITS INTENSITY. NONETHELESS, IT SEEMS MORE LIKELY THE BURN WOULD BE GENERALLY CIRCULAR. WITHOUT FURTHER INFORMATION THIS CANNOT BE PROPERLY ADDRESSED. ALL EQUIPMENT (TWO EACH LIGHTHOUSES, HEADLIGHTS AND FIBEROPTIC CABLES) WERE RETURNED TO US AND WERE EVALUATED BY OUR SERVICE AND REPAIR TECHNICIAN AS WELL AS AN ENGINEER. ALL EQUIPMENT WAS FOUND TO BE IN EXCELLENT WORKING CONDITION WITH NO ABNORMALITIES PRESENT.

Description of Event or Problem · 1

THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR STAGE II RIGHT BREAST RECONSTRUCTION WITH REMOVAL OF TISSUE EXPANDERS AND PLACEMENT OF SILICONE GEL IMPLANTS WITH LEFT BREAST REDUCTION FOR SYMMETRY PURPOSES. IT WAS NOTED DURING THE PROCEDURE THAT THE PATIENT HAD SUSTAINED A CRESCENT SHAPED FIRST DEGREE BURN ON THE SKIN AT THE RIGHT BREAST INCISION SITE. NO BLISTERS WERE SEEN. THE AREA WAS BANDAGED ALONG WITH THE ROUTINE BREAST DRESSINGS. IT WAS THOUGHT THAT THE HEADLIGHT THAT THE SURGEON WAS WEARING AND USING MAY HAVE CAUSED THE BURN BY HEATING THE TISSUE EXCESSIVELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUNOPTIC TECHNOLOGIES HEADLIGHT FSS SUNOPTIC TECHNOLOGIES C/O SERVICE AND TECH SUPPORT UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other