FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1913036
·
Received November 3, 2010
Report
- Report Number
- 2531779-2010-02080
- Event Type
- Malfunction
- Date Received
- November 3, 2010
- Date of Event
- October 5, 2010
- Report Date
- October 5, 2010
- Manufacturer
- ANIMAS CORP
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS AND THE EVAL REVEALED THAT THE KEYPAD WAS INTACT WITH NO LIFTING OR PEELING. ALL KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING. THE KEYPAD WAS REMOVED AND THE "UP/DOWN" KEY CONTACTS WERE OBSERVED TO BE MISALIGNED. THE "OK" KEY CONTACT WAS OBSERVED TO BE INVERTED.
Description of Event or Problem · 1
THE PT REPORTED THAT THE BUTTONS HAVE TO BE PRESSED MULTIPLE TIMES TO ELICIT A RESPONSE FROM THE KEYPAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |