FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 1913007 · Received November 18, 2010

Report

Report Number
3002158293-2010-01223
Event Type
Malfunction
Date Received
November 18, 2010
Date of Event
October 23, 2010
Report Date
November 16, 2010
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE BELT COMMUNICATION ERRORS WAS DUE TO A BROKEN TRUNK CABLE CONNECTOR. THE ROOT CAUSE OF THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE PT REC'D A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT OT REPORT THAT SHE IS GETTING A SERVICE CODE 204 AND A MESSAGE STATING THE DEVICE CANNOT BE USED. THE PT'S ELECTRODE BELT WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 22 YR