FDA Adverse Event
Malfunction
Summary report: N
4WEB ANTERIOR SPINE TRUSS SYSTEM - STAND ALONE
MDR report key: 19129344
·
Received April 17, 2024
Report
- Report Number
- 3009189869-2024-00007
- Event Type
- Malfunction
- Date Received
- April 17, 2024
- Date of Event
- March 5, 2024
- Report Date
- April 17, 2024
- Manufacturer
- 4WEB, INC.
- Product Code
- HWJ
- UDI-DI
- 00812998043903
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT HAS BEEN RETURNED TO THE MANUFACTURER YET. IT WAS REPORTED THAT THE PLACEMENT OF THE INTERBODY MADE ALL TRAJECTORIES CHALLENGING. THE PRODUCTION RECORDS WERE REVIEWED BASED ON THE PART AND LOT NUMBER INFORMATION PROVIDED. NO MANUFACTURING DEVIATIONS WERE IDENTIFIED THAT COULD HAVE CONTRIBUTED TO THIS EVENT. THE EXACT CAUSE OF THIS EVENT COULD NOT BE ESTABLISHED. THE EVENT WAS REPORTED TO HAVE OCCURRED IN AUSTRALIA.
Description of Event or Problem · 0
THE TIP OF THE STRAIGHT PUNCH AWL USED IN SURGERY BROKE WHEN SURGEON REMOVED THE AWL FROM THE CAGE. TIP OF AWL REMAINS IN PATIENT WITHIN THE INTEGRATED 4WEB CAGE. NO INJURY TO PATIENT WAS REPORTED. SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 539736 | 4WEB ANTERIOR SPINE TRUSS SYSTEM - STAND ALONE | STRAIGHT PUNCH AWL | HWJ | 4WEB, INC. | ASTS-SA-100006 | 30394 | 00812998043903 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |