FDA Adverse Event
Malfunction
Summary report: N
ADVANTA 2
MDR report key: 1912704
·
Received November 24, 2010
Report
- Report Number
- 3006697241-2010-00049
- Event Type
- Malfunction
- Date Received
- November 24, 2010
- Date of Event
- November 1, 2010
- Report Date
- November 1, 2010
- Manufacturer
- HILL-ROM DE MEXICO S DE RL DE CV
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACCOUNT HAS ISOLATED THE PENDANT AND THE ISSUE REMAINS. THE TECHNICIAN FOUND THE MOTOR CONTROL BOARD WAS NOT FUNCTIONING. REPAIRS HAVE NOT BEEN COMPLETED.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THAT THE TRENDELENBURG FUNCTION IS NOT WORKING ON THIS BED AND WHEN HE ATTEMPTS TO RAISE THE HI/LOW, THE HEAD END OF THE BED RISES BRIEFLY AND THE FOOT NEVER MOVES. NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA 2 | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM DE MEXICO S DE RL DE CV | 1190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |