FDA Adverse Event Injury Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1912700 · Received November 26, 2010

Report

Report Number
2531779-2010-02568
Event Type
Injury
Date Received
November 26, 2010
Date of Event
October 29, 2010
Report Date
October 29, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE PUMP WAS EVALUATED AND FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING INSULIN ACCURATELY. INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ISSUES. THE PUMP WAS PLACED IN THE LOCKED POSITION AND NO UNPROGRAMMED BOLUSES WERE DELIVERED OR RECORDED IN THE PUMP HISTORY. WITH THE PUMP LOCKED THE BOLUS SCREEN WAS UNABLE TO BE ACTIVATED. NO STICKING OR HYPERTENSIVE BUTTONS WERE OBSERVED.

Description of Event or Problem · 1

THE PT'S PARENTS REPORTED THAT THE PT EXPERIENCED HYPOGLYCEMIA (33MG/DL) WHILE AT SCHOOL. THE PT WAS TREATED BY THE SCHOOL NURSE WITH JUICE AND SODA. THE PUMP HISTORY INDICATES MULTIPLE BOLUSES WERE PERFORMED TOTALING 48 UNITS. THE PT'S PARENTS INDICATED THAT THE PUMP WAS LOCKED, BUT THE PT KNOWS HOW TO UNLOCK THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 9 YR Required Intervention