NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Report
- Report Number
- 1222780-2010-00177
- Event Type
- Injury
- Date Received
- November 24, 2010
- Date of Event
- October 25, 2010
- Report Date
- October 25, 2010
- Manufacturer
- HOLOGIC
- Product Code
- MNB
- PMA / PMN Number
- P010013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
SERIAL NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT. SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. THE DEVICE IS NOT BEING RETURNED, THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. BASED ON THE INFO OBTAINED TO DATE, NO DIRECT CORRELATION CAN BE MADE BETWEEN THE REPORTED EVENT AND THE NOVASURE SYSTEM. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGEMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).
THE PHYSICIAN ATTEMPTED A NOVASURE ENDOMETRIAL ABLATION, BUT THE CAVITY INTEGRITY ASSESSMENT (CIA) TEST WAS UNSUCCESSFUL. THE PHYSICIAN BELIEVED THERE TO BE A UTERINE PERFORATION AND ABANDONED THE PROCEDURE. NO TREATMENT WAS NEEDED AND THE PT WAS GIVEN PROPHYLACTIC ANTIBIOTICS AND DISCHARGED HOME. A DILATATION, BIOPSY, AND SOUNDING WITH A METAL SOUND (NOT A HOLOGIC DEVICE) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NOVASURE IMPEDANCE CONTROLLED EA SYSTEM | MNB | HOLOGIC | NS2000 | 09K20RD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Other | RADIO FREQUENCY CONTROLLER - SERIAL NUMBER UNK |