FDA Adverse Event Injury Summary report: N

WELCH ALLYN CONNEX 6800 VITAL SIGNS MONITOR

MDR report key: 19126077 · Received April 16, 2024

Report

Report Number
MW5153907
Event Type
Injury
Date Received
April 16, 2024
Report Date
April 12, 2024
Manufacturer
WELCH ALLYN, INC.
Product Code
MHX
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

WELCH ALLYN CONNEX 6800 VITAL SIGNS MONITOR FELL OFF THE WALL MOUNTED ARM AND ADAPTER BRACKET. THIS IS THE 3RD TIME IN 4 MONTHS THAT THIS HAS OCCURRED IN ONE PATIENT CARE AREA (FLOOR A1) WHICH HAS 28 OF THESE WALL MOUNTED MONITORS. IT APPEARS THAT OVER TIME, THROUGH NORMAL MANIPULATION OF THE MONITOR, THE THUMBSCREW, THAT SECURES THE MONITOR TO THE BRACKET, LOOSENS AND THE BRACKET BECOMES BENT. THIS ALLOWS THE MONITOR TO RELEASE FROM THE BRACKET AND FALL. THESE ARE MOUNTED AT THE HEAD OF THE PATIENT BED AND CAN FALL ON PATIENTS AND CLINICAL STAFF. ONE OF THE 3 THAT HAS RECENTLY FALLEN, INJURED A NURSE. IN ANOTHER INPATIENT AREA (FLOOR A2), THE MOUNTING BRACKETS ON 6 OF THE 30 MONITORS HAD TO BE REPLACED DURING THE MONTH OF (B)(6) 2024 TO PREVENT THESE MONITORS FROM FALLING. REFERENCE REPORTS: MW5153905, MW5153906.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539378 WELCH ALLYN CONNEX 6800 VITAL SIGNS MONITOR MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) MHX WELCH ALLYN, INC.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention