FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1912569 · Received December 2, 2010

Report

Report Number
1423500-2010-06449
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
October 30, 2010
Report Date
November 20, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE RITE (RETURN INSTRUMENT TEST/EVALUATION) TEST WAS PERFORMED WHEN THE DEVICE WAS RETURNED TO THE BAXTER (B)(4) FACILITY FOR EVALUATION. THE DEVICE PASSED THE HOMECHOICE RITE FUNCTIONAL TEST AND THE HOMECHOICE RITE ELECTRICAL TEST. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION IN THE EVENT LOG WHICH OCCURRED ON (B)(6) 2010 AT 05:00:18 WITH DRAIN VOLUME OF 2776ML DURING CYCLE 7. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1