EVIS EXERA III VIDEO SYSTEM CENTER
Report
- Report Number
- 3002808148-2024-03570
- Event Type
- Malfunction
- Date Received
- April 17, 2024
- Date of Event
- April 2, 2024
- Report Date
- May 7, 2024
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FET
- UDI-DI
- 04953170298622
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. CORRECTED FIELDS: E1 (ESTABLISHMENT NAME). ADDITIONAL INFORMATION ADDED TO FIELDS: H3 AND H6. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 9 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S REPORTABLE MALFUNCTION WAS NOT CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION, AND SINCE THE DEVICE MALFUNCTION WAS NOT CONFIRMED DURING EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
F24149684-1 : [FINAL EVALUATION RESULT] EVALUATION DATE: 2024/04/11 EVALUATOR: (B)(6). PURPOSE OF PROCEDURE: UNKNOWN EVENT:THE IMAGE COLOR WAS ABNORMAL DUE TO FAULTY DP BOARD. PAE JUDGMENT: PAE-NO INVESTIGATION REQUIRED: NO INVESTIGATION REQUIRED TIER CLASSIFICATION: 3 UNIVERSAL FAILURE CODE: CVP1143G CONTAINMENT MEASURES: NO (AT THIS TIME, NO FACTS HAVE BEEN CONFIRMED TO SUGGEST THAT THE REPORTED EVENT MAY EXPAND.) F24149684-2 : [FINAL EVALUATION RESULT] EVALUATION DATE: 2024/04/15 EVALUATOR: (B)(6). PURPOSE OF PROCEDURE: UNKNOWN EVENT:THE IMAGE COLOR WAS ABNORMAL DUE TO FAULTY DP BOARD. PAE JUDGMENT: PAE-NO INVESTIGATION REQUIRED: NO INVESTIGATION REQUIRED TIER CLASSIFICATION: 3 UNIVERSAL FAILURE CODE: CVP1143G EVENT:FRONT PANEL SWITCHES NOT FUNCTIONING PAE JUDGMENT: PAE-YES INVESTIGATION REQUIRED: INVESTIGATION REQUIRED TIER CLASSIFICATION: 2 UNIVERSAL FAILURE CODE: CVP3132 CONTAINMENT MEASURES: NO (AT THIS TIME, NO FACTS HAVE BEEN CONFIRMED TO SUGGEST THAT THE REPORTED EVENT MAY EXPAND.)
IT WAS REPORTED THE VIDEO SYSTEM CENTER FRONT PANEL BUTTONS DIDN'T WORK. THERE WERE NO REPORTS OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1024829 | EVIS EXERA III VIDEO SYSTEM CENTER | VIDEO SYSTEM CENTER | FET | SHIRAKAWA OLYMPUS CO., LTD. | CV-190 | 04953170298622 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |