FDA Adverse Event Malfunction Summary report: N

BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML

MDR report key: 19125363 · Received April 17, 2024

Report

Report Number
1119779-2024-00298
Event Type
Malfunction
Date Received
April 17, 2024
Date of Event
March 27, 2024
Report Date
September 9, 2024
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
MDB
UDI-DI
30382902451229
PMA / PMN Number
K974883
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: MATERIAL 245122 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE FILLED, CAPPED, TORQUED AND THEN LABELED BY MACHINE PER STANDARD OPERATING PROCEDURE (SOP). THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCHES 3011178 AND 3145748 WAS SATISFACTORY PER INTERNAL PROCEDURES. FORMULATION, FILLING, TORQUING, PACKAGING PROCESSES WERE WITHIN SPECIFICATIONS. IN PROCESS CHECKS WERE PERFORMED DURING MANUFACTURING AT DESIGNATED INTERVALS PER PROCEDURES. CHECKS FOR FILL VOLUME WERE COMPLETE AND WITHIN SPECIFICATIONS PER PROCEDURES AND CHECKS FOR TORQUE CONFIRMED THAT THE CAPS WERE TIGHTENED TO THE VALIDATED SPECIFICATIONS PER INTERNAL PROCEDURE. QC INSPECTION AND TESTING WERE SATISFACTORY AT TIME OF RELEASE. THE COMPLAINT HISTORY WAS REVIEWED, AND OTHER COMPLAINTS HAVE BEEN MADE ON BATCH 3011178 AND 3145748 FOR LEAKING AND NO OTHER COMPLAINTS WERE TAKEN ON BATCH 3145748 FOR MISSING LABELS. RETENTION SAMPLES (100 PER BATCH) WERE AVAILABLE FOR INSPECTION. THERE WERE NO MISSING LABEL DEFECTS, LOW FILLS, OR LEAKING OBSERVED IN THE RETENTION SAMPLES. ONE PHOTO AND A COLLAGE OF THREE PHOTOS WERE SUBMITTED TO REVIEW FOR THIS COMPLAINT. THE FIRST PHOTO SHOWS A TUBE TRAY WITH FIVE TUBES. THE TWO TUBES WITH BATCH INFORMATION ARE FROM BATCH 3145748 WITH EXPIRATION 2024-11-24. THE THIRD AND FOURTH TUBES FROM THE LEFT ARE MISSING LABELS. THE TUBE THIRD FROM THE RIGHT ALSO HAS A LOW FILL. THE OTHER FOUR TUBES CANNOT BE ASSESSED FOR FILL VOLUME USING THIS PHOTO. THE PHOTO COLLAGE CONTAINS THREE ADDITIONAL PHOTOS. THE FIRST PHOTO IN THE COLLAGE SHOWS TWO TUBES. THE TUBE ON THE LEFT IS MISSING A LABEL AND THE TUBE ON THE RIGHT IS FROM BATCH 3145748 AND HAS A LOW FILL. THE SECOND PHOTO IN THE COLLAGE IS A TUBE FROM BATCH 3011178 WITH EXPIRATION 2024-07-07. THE TUBE APPEARS TO HAVE A LOW FILL. WHILE LEAKING CANNOT BE ASSESSED IN THE IMAGE, THE LOW FILL IS INDICATIVE OF LEAKING. THE FINAL PHOTO IN THE COLLAGE IS A TUBE FROM BATCH 3145748 WITH EXPIRATION 2024-11-24. THE TUBE IN THIS IMAGE IS EMPTY. NO RETURNS WERE RECEIVED TO ASSIST IN THE INVESTIGATION OF THIS COMPLAINT. THIS COMPLAINT CAN BE CONFIRMED FOR MISSING LABELS, LOW FILL, AND LOOSE CAP. NO TREND IDENTIFIED. BD WILL CONTINUE TO TREND COMPLAINTS.

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER FACILITY NAME: (B)(6). H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML THE PRODUCT LABEL WAS MISSING ON TWO (2) TUBES. THERE WAS NO REPORT OF IMPACT TO THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML THE PRODUCT LABEL WAS MISSING ON TWO (2) TUBES. THERE WAS NO REPORT OF IMPACT TO THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1023799 BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML SYSTEM, BLOOD CULTURING MDB BECTON, DICKINSON & CO. (SPARKS) 3145748 30382902451229

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown