FDA Adverse Event Malfunction Summary report: N

ECHELON* 60 ENDOPATH

MDR report key: 1912480 · Received December 2, 2010

Report

Report Number
3005075853-2010-06804
Event Type
Malfunction
Date Received
December 2, 2010
Report Date
November 9, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT ONE LONG60 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE OPENED WITHOUT ANY DIFFICULTIES NOTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLECTOMY PROCEDURE, THE CLIPS WERE MALFORMED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE. NO MORE DETAILS AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY PROCEDURE, AFTER THE DEVICE COMPLETED THE FIRING THE DEVICE WAS HANDED TO THE BACK TABLE AND THEY COULD NOT GET THE DEVICE OPEN TO RELOAD THE DEVICE. THE DEVICE WAS NOT ON TISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON* 60 ENDOPATH STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1