FDA Adverse Event Malfunction Summary report: N

REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE

MDR report key: 1912417 · Received December 2, 2010

Report

Report Number
6000001-2010-05360
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
October 18, 2010
Report Date
October 18, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF ?FAILURE CODE 812.02.?? (B)(4).

Additional Manufacturer Narrative · 1

THE REPORTED CONDITION OF FAILURE CODE 812.02 WAS CONFIRMED AND DUPLICATED. THE ROOT CAUSE HAS NOT BEEN DETERMINED AT THIS TIME. NO REPAIRS HAVE BEEN PERFORMED AT THIS TIME SINCE THIS IS A BAXTER OWNED DEVICE. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4)

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE PUMP WITH FAILURE CODE 812.02. IT WAS NOT SPECIFIED WHEN REPORTED CONDITION OCCURRED. THERE WAS NO DOCUMENTED PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. DURING REVIEW OF THE EVENT HISTORY BY BAXTER IT WAS DETERMINED THAT THE REPORTED CONDITION OCCURRED DURING DELIVERY CAUSING AN INTERRUPTION OF DELIVERY. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90 CATEGORIZED AS REMEDIATED.