FDA Adverse Event
Injury
Summary report: N
ACHIEVA 1.5T NOVA
MDR report key: 1912387
·
Received November 16, 2010
Report
- Report Number
- 3003768277-2010-00309
- Event Type
- Injury
- Date Received
- November 16, 2010
- Date of Event
- October 19, 2010
- Report Date
- October 19, 2010
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- LNH
- PMA / PMN Number
- K052013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(METHOD, RESULTS, CONCLUSIONS) - THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED, A F/U REPORT WILL BE SENT TO THE FDA. (B)(4).
Description of Event or Problem · 1
THE PATIENT WAS SCANNED IN THE HEAD FIRST POSITION WITH THE SENSE HEAD COIL. THE CABLES WERE BESIDES THE RIGHT ARM OF THE PATIENT; NO PROTECTIVE MEASURES WERE TAKEN TO SEPARATE THE CABLE FROM THE PATIENT'S ARM. AFTER THE EXAM, A SECOND DEGREE BURN WITH A SIZE OF APPROX 1X3 WAS FOUND ON THE RIGHT ARM, BETWEEN THE HAND AND THE ELBOW. INVESTIGATION IS ONGOING, FURTHER DETAILS WILL FOLLOW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACHIEVA 1.5T NOVA | LNH (MAGNETIC RESONANCE DIAGNOSTIC DEVICE) | LNH | PHILIPS MEDICAL SYSTEMS | 781196 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Other |