FDA Adverse Event
Injury
Summary report: N
INTRAMATIC LUX3 CONTRA ANGLE 25LHA
MDR report key: 1912367
·
Received November 19, 2010
Report
- Report Number
- 3003637274-2010-00044
- Event Type
- Injury
- Date Received
- November 19, 2010
- Date of Event
- October 28, 2010
- Report Date
- November 3, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GOT HOT AND PT RECEIVED A SMALL BURN ON CHEEK. DURING PRODUCT EVAL WAS FOUND THAT BEARINGS ARE GRITTY AND FALLING APART. THIS CAUSE THE HANDPIECE TO HEAT UP. HANDPIECE DID NOT GET A REPAIR SINCE 05/2004. (B)(4). KAVO DENTAL (B)(4) (THE MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4) (THE IMPORTER).
Description of Event or Problem · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GOT HOT AND PT RECEIVED A SMALL BURN ON CHEEK. TWO PTS HAVE BEEN BURNED IN THE SAME WAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTRAMATIC LUX3 CONTRA ANGLE 25LHA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LHA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |