FDA Adverse Event Injury Summary report: N

INTRAMATIC LUX3 CONTRA ANGLE 25LHA

MDR report key: 1912367 · Received November 19, 2010

Report

Report Number
3003637274-2010-00044
Event Type
Injury
Date Received
November 19, 2010
Date of Event
October 28, 2010
Report Date
November 3, 2010
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GOT HOT AND PT RECEIVED A SMALL BURN ON CHEEK. DURING PRODUCT EVAL WAS FOUND THAT BEARINGS ARE GRITTY AND FALLING APART. THIS CAUSE THE HANDPIECE TO HEAT UP. HANDPIECE DID NOT GET A REPAIR SINCE 05/2004. (B)(4). KAVO DENTAL (B)(4) (THE MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4) (THE IMPORTER).

Description of Event or Problem · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE HANDPIECE GOT HOT AND PT RECEIVED A SMALL BURN ON CHEEK. TWO PTS HAVE BEEN BURNED IN THE SAME WAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTRAMATIC LUX3 CONTRA ANGLE 25LHA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LHA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention