FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONRA ANGLE 25LPA
MDR report key: 1912366
·
Received November 19, 2010
Report
- Report Number
- 3003637274-2010-00046
- Event Type
- Injury
- Date Received
- November 19, 2010
- Date of Event
- October 12, 2010
- Report Date
- October 21, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
WHILE WORKING ON THE TOOTH# 12, THE PT'S LOWER LIP WAS BURNED. THE BURN AREA WAS THE SIZE OF HANDPIECE HEAD. THE PT WAS GIVEN SOME MEDICATED VASELINE. THE INJURY DOES HEAL WELL. DURING PRODUCT EVAL, LOW DEBRIS LEVEL WAS FOUND IN THE HANDPIECE. THE BEARINGS WERE GRITTY. THE GRINDING BEARINGS LEAD TO HEAT UP THE HEAD OF THE HANDPIECE. (B)(4). KAVO DENTAL (B)(4) (THE MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4) (THE IMPORTER).
Description of Event or Problem · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PT'S LOWER LIP WAS BURNED BY THE HEAD OF THE HANDPIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |