FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONRA ANGLE 25LPA

MDR report key: 1912366 · Received November 19, 2010

Report

Report Number
3003637274-2010-00046
Event Type
Injury
Date Received
November 19, 2010
Date of Event
October 12, 2010
Report Date
October 21, 2010
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WHILE WORKING ON THE TOOTH# 12, THE PT'S LOWER LIP WAS BURNED. THE BURN AREA WAS THE SIZE OF HANDPIECE HEAD. THE PT WAS GIVEN SOME MEDICATED VASELINE. THE INJURY DOES HEAL WELL. DURING PRODUCT EVAL, LOW DEBRIS LEVEL WAS FOUND IN THE HANDPIECE. THE BEARINGS WERE GRITTY. THE GRINDING BEARINGS LEAD TO HEAT UP THE HEAD OF THE HANDPIECE. (B)(4). KAVO DENTAL (B)(4) (THE MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF (B)(4) (THE IMPORTER).

Description of Event or Problem · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PT'S LOWER LIP WAS BURNED BY THE HEAD OF THE HANDPIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONRA ANGLE 25LPA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention