FDA Adverse Event Malfunction Summary report: N

SMARTTOUCH

MDR report key: 19123230 · Received April 17, 2024

Report

Report Number
1000165971-2024-00326
Event Type
Malfunction
Date Received
April 17, 2024
Date of Event
March 21, 2024
Report Date
July 18, 2024
Manufacturer
MICROPORT CRM S.R.L.
Product Code
MRM
PMA / PMN Number
P980049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION LED TO THE FOLLOWING OBSERVATIONS: - THE POP-UPS AND/OR PROGRAMMING MENUS AND/OR DROPDOWN LIST ARE NOT VISIBLE BY THE USER. INDEED, THEY ARE PRESENT BUT DISPLAYED BEHIND THE MAIN SCREEN. THEREFORE, THE USER IS NOT ABLE TO CHOOSE/CLICK ON THE APPROPRIATE ANSWER/PARAMETER AND THIS EXPLAINS THE REPORTED FREEZE DURING THE PROGRAMMER SESSION. - THIS ISSUE HAS BEEN REPRODUCED BY THE R&D TEAM. - THE MOST LIKELY HYPOTHESIS IS A PROGRAMMER SOFTWARE ISSUE, AND IT WILL BE CORRECTED WITH THE NEXT SMARTVIEW VERSION. - THE COMPLAINT IS RECORDED FOR TRENDING PURPOSES.

Description of Event or Problem · 0

OA PARADYM RF DR WAS "INTERROGATED" WITH SV 3.16 IN THE PILOT CENTER (B)(6). AFTER THE MANUAL RV THRESHOLD TEST, IT WAS NO LONGER POSSIBLE TO SELECT THE RA THRESHOLD TEST OR ANY OTHER FUNCTION. THE APPLICATION FROZE AND I HAD TO REMOVE THE TABLET BATTERY. I RAN THE INTEGRATION AGAIN WITH ANOTHER TABLET RUNNING 3.14 AND ALL TESTS WORKED FINE.

Description of Event or Problem · 0

OA PARADYM RF DR WAS INTEROGATED WITH SV 3.16 IN THE PILOT CENTER AKH WIEN. AFTER THE MANUAL RV THRESHOLD TEST, IT WAS NO LONGER POSSIBLE TO SELECT THE RA THRESHOLD TEST OR ANY OTHER FUNCTION. THE APPLICATION FROZE AND I HAD TO REMOVE THE TABLET BATTERY. I RAN THE INTEGRATION AGAIN WITH ANOTHER TABLET RUNNING 3.14 AND ALL TESTS WORKED FINE.

Description of Event or Problem · 0

A PARADYM RF DR WAS INTEROGATED WITH SV 3.16 IN THE PILOT CENTER AKH WIEN. AFTER THE MANUAL RV THRESHOLD TEST, IT WAS NO LONGER POSSIBLE TO SELECT THE RA THRESHOLD TEST OR ANY OTHER FUNCTION. THE APPLICATION FROZE AND I HAD TO REMOVE THE TABLET BATTERY. I RAN THE INTEGRATION AGAIN WITH ANOTHER TABLET RUNNING 3.14 AND ALL TESTS WORKED FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1914592 SMARTTOUCH DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER MRM MICROPORT CRM S.R.L. SMARTTOUCH SOFTWARES MODULES
2320080 SMARTTOUCH DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER MRM MICROPORT CRM S.R.L. SMARTTOUCH SOFTWARES MODULES

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown