FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36CM W ERG HANDLE

MDR report key: 1912282 · Received December 2, 2010

Report

Report Number
3005075853-2010-06781
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
October 28, 2010
Report Date
November 11, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED WITH THE TISSUE PAD PARTIALLY DETACHED BUT WITH EVIDENCE OF THE TISSUE PAD MATERIAL IN THE GROOVE SECTION OF THE CLAMP ARM. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM; AND ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. BOTH CONDITIONS CAN RESULT IN PROBABLE DAMAGE TO THE INSTRUMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY PROCEDURE THE TISSUE PAD MELTED AND THEN FELL OFF. THE TISSUE PAD FELL INTO THE PATIENT, BUT WAS RETRIEVED THROUGH THE TROCAR. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36CM W ERG HANDLE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE