MICROSTAAR INJECTOR
Report
- Report Number
- 2023826-2010-01169
- Event Type
- Malfunction
- Date Received
- November 23, 2010
- Date of Event
- October 26, 2010
- Report Date
- October 27, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- KYB
- PMA / PMN Number
- K954600
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
(B)(4). THE PRODUCT PERTAINING TO THIS COMPLAINT WAS NOT RETURNED FOR EVAL. HOWEVER, THE LENS WAS RETURNED AND VISUAL INSPECTION FOUND HAPTIC TORN OFF AND MISSING. LENS WAS RETURNED IN LENS CASE. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. EVAL CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE REPORTER STATED THE SURGEON INSERTED AN AQ2015A 22.5 DIOPTER LENS. THE LENS TORE ON INSERTION INTO THE EYE. LENS WAS REMOVED WITH NO PT INJURY. NO WIDENED INCISION AND NO SUTURE. BACKUP LENS WAS IMPLANTED USING A NEW INJECTOR. THE REPORTER STATED THE LENS WAS DAMAGED BY THE INJECTOR. NO LOT NUMBER WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSTAAR INJECTOR | INTRAOCULAR LENS GUIDE | KYB | STAAR SURGICAL CO. | MSI-PM | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | LENS: MODEL AQ2015A -(B)(4)| VISCOELASTIC: -STAARVISC II -LOT # UNK| CARTRIDGE: MODEL AQ CARTRIDGE -LOT # UNK |