FDA Adverse Event Malfunction Summary report: N

MICROSTAAR INJECTOR CARTRIDGE

MDR report key: 1912257 · Received November 23, 2010

Report

Report Number
2023826-2010-01199
Event Type
Malfunction
Date Received
November 23, 2010
Date of Event
October 26, 2010
Report Date
November 11, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
KYB
PMA / PMN Number
K954600
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT PERTAINING TO THIS COMPLAINT WAS NOT RETURNED FOR EVAL. HOWEVER, THE LENS WAS RETURNED AND VISUAL INSPECTION FOUND PIECES OF THE OPTIC AND A HAPTIC TORN OFF AND MISSING. EVAL: CARTRIDGE LOT NUMBER SEARCH. RESULTS: CARTRIDGE LOT NUMBER SEARCH WAS PERFORMED AND FOUR SIMILAR COMPLAINTS WERE FOUND WITHIN THE SAME LOT NUMBER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MFG, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED A 22.5 DIOPTER CQ2015A COLLAMER ASPHERIC THREE PIECE LENS. THE LENS WAS REMOVED DUE TO BROKEN HAPTIC. THERE WAS NO PT INJURY. ANOTHER SAME MODEL AND SIZE LENS WAS IMPLANTED. REPORTER STATED THE CARTRIDGE DAMAGED THE LENS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROSTAAR INJECTOR CARTRIDGE INTRAOCULAR LENS GUIDE KYB STAAR SURGICAL CO. CQ CARTRIDGE-FP 1254557

Patients

Seq Age Sex Outcome Treatment
1 67 YR LENS: MODEL CQ2015A, (B)(4)| INJECTOR: MODEL -MSI-PM, LOT # UNK