OT PING METER
Report
- Report Number
- 2939301-2010-10427
- Event Type
- Malfunction
- Date Received
- December 2, 2010
- Report Date
- November 16, 2010
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT-12/21/2010. THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE FAILED TESTING. THE METER WAS FOUND TO HAVE BATTERY CONTACT CORRODED. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510(K)# IS K082590.
CUSTOMER REPORTED THAT THE DEVICE HAD THE WRENCH ICON ILLUMINATED. PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT SHOWED THE SERVICE WRENCH AND THE "CALL SERVICE" MESSAGE. THE DEVICE LOCKED UP AND WAS NOT AVAILABLE TO PROVIDE DEFIBRILLATION THERAPY. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE METER REVERTING TO SETUP MODE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3062505 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |