PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2010-03050
- Event Type
- Injury
- Date Received
- December 1, 2010
- Date of Event
- November 8, 2010
- Report Date
- November 10, 2010
- Manufacturer
- AV-REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS. SUBSEQUENT INFORMATION REVEALED THAT THE DEVICE IS NOT AVAILABLE FOR EVALUATION. THE LOT NUMBER WAS NOT IDENTIFIED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.(B)(4).
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. (B)(4) - MORBID OBESITY. (B)(4): DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. DEVICE #1 - PROGLIDE (PART #12673-03; LOT #UNK), ARE BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, THE DEVICE COULD NOT BE INSERTED THROUGH THE SHEATH. THE STARCLOSE SE SYSTEM WAS REMOVED AND MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, A KINK WAS FOUND IN THE DISTAL END OF THE SHEATH. THE GUIDE WIRE WAS DIFFICULT TO ADVANCE INTO THE PROGLIDE AND AFTER CONTINUALLY TRYING TO ADVANCE THE GUIDE WIRE, THE GUIDE WIRE PUNCTURING A HOLE THROUGH THE SHEATH CREATING ITS OWN LUMEN. THE TIP OF THE PROGLIDE FOLDED BACK INTO THE PATIENT'S ILIAC, POSSIBLY BECAUSE OF STENOSIS. THE DEVICE WAS REMOVED AND A SECOND PROGLIDE DEVICE WAS USED; HOWEVER, HEMOSTASIS WAS NOT ACHIEVED. HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION FOR APPROXIMATELY 15 MINUTES. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-REDWOOD CITY |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | DEVICE #1: PROGLIDE (PART #12673-03, LOT #UNK) |