FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-01621
- Event Type
- Malfunction
- Date Received
- December 1, 2010
- Date of Event
- November 5, 2010
- Report Date
- December 2, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
THIS IS A FINAL REPORT. CUSTOMER'S METER (B)(4) WAS RETURNED AND TESTED WITH TEST STRIPS LOT # 1067323. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED READINGS WERE FOUND IN THE DEVICE'S INTERNAL MEMORY LOG WITHIN 10 MINUTES. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 1.1 MMOL/L TO 27.8 MMOL/L. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 27.8 MMOL/L. A BLOOD GLUCOSE READING ABOVE 27.8 MMOL/L WILL READ AS A "HI" IN THE ADC METER.
(B)(4).
A HEALTH CARE PROFESSIONAL REPORTED THAT A CUSTOMER'S METER WAS RECEIVING ERRATIC READINGS. CALLER REPORTED RECEIVING READINGS OF "HI" (GREATER THAN 27.8 MMOL/L), "HI" (GREATER THAN 27.8 MMOL/L) AND 1.6 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "D" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
THE REPORT RECEIVED FROM THE FIELD STATES THAT WHEN ATTEMPTING TO CROSS THE LESION WITH THE STENT DELIVERY SYSTEM (SDS) THE STENT DISLODGED. THE AGE AND GENDER OF THE PATIENT ARE UNKNOWN. THE PROCEDURE WAS FOR TREATMENT OF A STENOSIS IN A PROXIMAL VEIN GRAFT. THE CHARACTERISTICS OF THE LESION ARE UNKNOWN. AS PER THE REPORT, THE PHYSICIAN DID NOT KNOW IF THE STENT WAS STILL ON THE BALLOON WHILE TRYING TO CROSS THE LESION. HE THEN PULLED BACK AND CHECKED THE STENT. THE PHYSICIAN THOUGHT THE STENT FELT STRANGE ON THE BALLOON CATHETER. A SECOND ATTEMPT WAS MADE TO CROSS THE LESION AND THIS IS WHEN THE STENT CAME OFF THE BALLOON. HE WENT IN WITH A 2.0 BALLOON AND SAVED THE STENT IN THE PROXIMAL VEIN GRAFT. THE STENT WAS SUCCESSFULLY DEPLOYED. NO PATIENT INJURY. THE PHYSICIAN'S CONCERN WAS WHY THIS HAPPENED - POSSIBLE STRUT BEING CAUGHT ON THE STENOSIS OR CAUGHT ON THE GUIDE. THE GUIDE USED WAS MEDTRONIC LAUNCHER JR 4 WITH OUTSIDE HOLES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1067323 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |