FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1911913
·
Received November 24, 2010
Report
- Report Number
- 2531779-2010-02528
- Event Type
- Malfunction
- Date Received
- November 24, 2010
- Date of Event
- October 27, 2010
- Report Date
- October 27, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS AND THE EVALUATION REVEALED THAT THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING. THE "DOWN" KEYPAD BUTTON WAS INTERMITTENTLY RESPONDING. THE "UP" AND "OK" BUTTONS WERE RESPONSIVE TO USER INPUTS AND DID CLICK AND SPRING BACK NORMALLY. THE KEYPAD WAS REMOVED AND THE "DOWN" KEY CONTACT WAS MISALIGNED. ADDITIONALLY, THE "DOWN" KEY CONTACT BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK. THERE WAS ALSO EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THE PT REPORTED THAT THE "DOWN" ARROW BUTTON WAS SLOW TO RESPOND ON THE KEYPAD. THE "DOWN" ARROW BUTTON DID NOT SPRING BACK. THE REPORTER DENIES DAMAGE TO THE KEYPAD OR EXPOSURE TO MOISTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |